AGFA DX-D 400 User Manual page 143

Hide thumbs Also See for DX-D 400:
Table of Contents

Advertisement

6.5
X-RAY HANDSWITCH
OFF
Prep
Exp
6.6
HEAT UNITS
6.7
EXPOSURE COUNTERS
0232F EN 20191213
Radiographic exposures can also be initiated with the X-ray Handswitch which
is connected to the Control Console.
The X-ray Handswitch button has three positions: "Off", "Preparation", and
"X-ray Exposure", which operate in the same way that "Prep" and "Exp"
push-buttons on the Control Console.
Press the Handswitch half-way for "Prep" and fully for "Exp".
This X-ray Generator is equipped with a Heat Unit Calculator. During
exposures, the Heat Units are calculated and totalled.
To view the remaining Heat Units, press the "On" push-button. The kVp Display
shows the percentage of Heat Units that remain preceded by the letter "H". For
example, a display of "H75" would indicate that 75% of Heat Units capacity of
the X-ray Tube remains. "H -- -- " indicates that all the capacity remains. The kVp
Display reverts to its normal function after releasing the "On" push-button.
The operator can read the number of exposures made by the Generator, as
indicated below:
RAD EXPOSURES
IN TUBE-1
RAD EXPOSURES
IN TUBE-2
The number of exposures is shown on the kVp and mAs Displays, up to a
maximum of 999,999 exposures.
kV
Press and hold
Press and hold
= 123.456 exposures
mAs
DX-D 400
User Manual
and press once
and press once
133