Table 5-1 (cont.)
Error Codes
ERROR
DESCRIPTION
E41 to E46
System failure related to Dosimeter.
Collimator Error. A failure on the Automatic Collimator has
been detected (blades are full open or in movement during
E48
exposure, etc. )
E50
Exposure has been aborted by the Operator.
E51 to E93
System failure related to High Speed Rotor Controller.
E95
Exposure aborted by the Rapid Termination.
X-ray order not acknowledged.
E97
The Receptor (Digital Detector) did not receive or
acknowledge the X-ray order sent by the Generator.
E98
Service Mode Active.
0232F EN 20191213
WHAT TO DO
Touch the "Reset" button.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
If the equipment remains inoperative turn it OFF and call Field Service
Touch the "Reset" button.
Select the correct Ion Chamber or modify parameters.
Repeat the exposure.
If the equipment remains inoperative, turn it OFF and call Field Service.
Touch the "Reset" button.
Repeat the exposure.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
Touch the "Reset" button and call Field Service. This error does not
inhibit normal operation.