DX-D 400
User Manual
.
Note
112
A virtual keyboard appears on the screen. Write the name of the new Region,
View or Projection and click on "Enter". The new label of the Region, View or
Projection is located in the last position of the list.
Now, if what it was created was a Region, the operator will have to create
(assign) Views and Projections to that new Region, for that, the program
requires to create at least one View for that Region and after that, create the
Projections for that View (if required). Repeat the process to create and name
the new Views and Projections.
The warning message "APR technique not defined " is displayed on the "Edit"
menu. Set the Patient Size, Radiographic Parameters and Selections
(Workstation and AEC if available) related to that Region / View / Projection and
touch the button "Set Technique" .
If an APR technique is to be stored with AEC parameters, a
suitable back-up Time (or mAs) MUST be stored by the operator
for this APR technique.
SET TECHNIQUE: Touch this button to assign the selected Patient Size,
Radiographic Parameters and Selections (Workstation and AEC if available) to
the new or existing Region / View / Projection. A pop message requires a
confirmation.
MODIFY: Select a label (Region, View or Projection), then touch this button to
modify the label name using the virtual keyboard that appears on the screen.
Write the new name and click on "Enter".
DELETE: To delete a Region, View or Projection select the respective label and
touch this button. When a Region is deleted, all the labels and APR techniques
(Views and Projections) related to this Region are deleted. When a View is
deleted, all the labels and APR techniques (Projections) related to this View are
deleted.
0232F EN 20191213