4.6.7
USING AND MAINTAINING THE DIGITAL DETECTOR
.
Note
0232F EN 20191213
Before Exposure, check the equipment daily and confirm that it works properly.
The action of the Air-Conditioning or Heating may produce condensation in the
equipment, wait until the condensation evaporates before performing an
exposure. As a general rule, raise or lower the room temperature gradually to
avoid condensation.
During exposure, do not use the Detector near devices generating a strong
magnetic field.
For Wireless Detectors, do not cover the IR Data Port with hands or other parts
of the body and do not use the selected frequency channel (2.4 GHz band) for
other wireless devices.
After every examination, wipe with a cloth slightly damped the patient contact
surfaces as well as the handle and Grid with disinfectants such as ethanol. For
cleaning, wipe with a cloth damped in neutral detergent.
For further information on the Digital Detector Handling and
Maintenance, refer to the Digital Detector manuals.