LOCK
Table 4-1
Error Codes
ERROR
DESCRIPTION
Exposure" or/and "Preparation" orders are
"E06"
activated during power-up.
"E10"
Failure in memory Integrated Circuit.
"rAN Err"
Failure in Microcontroller.
0232F EN 20191213
TRANSVERSE MOVEMENT: Activate this button to allow transverse
movement of Tube-Collimator Assembly. Release button to lock in position.
This movement is provided with a detent when the Tube-Collimator Assembly
is centered with respect to the RAD Table Receptor ("LOCK" indicator lits).
Press this button again to continue movement.
ALL MOVEMENTS: Activate this button to allow Horizontal--Vertical--Rotation
movement at the same time. Release button to lock in position.
Operating this button disables the corresponding detents.
DIGITAL CONTROL PANEL ERROR CODES: Error codes indicate the
potential cause of a system failure. They appear on the Digital Control Panel
Display and will enable the operator to indirectly convey the possible source of
error to service personnel. This may prevent the need for a service call or enable
service personnel to anticipate corrective actions prior to arriving on site.
WHAT TO DO
Turn the Generator OFF,
check the proper external cable connections and then turn
the Generator ON. If the equipment remains inoperative,
turn it OFF and call Field Service.
Turn the System OFF / ON. If the error remains, turn it
OFF and call Field Service.
Turn the System OFF / ON. If the error remains, turn it
OFF and call Field Service.
DX-D 400
User Manual
57