DX-D 400
User Manual
5.1
RADIOGRAPHY AND GENERAL CONTROLS
5.1.1
POWER ON / OFF
.
Note
5.1.2
SERVICE MODE
96
ON: The Generator and the Console are turned ON by pressing and holding for
two seconds the "ON / OFF Switch" of the Console Housing. This starts the
application and the last parameters and workstation used are automatically
selected.
OFF: The Generator and Console are turned OFF by touching this button on the
Touch Screen.
For Generators equipped with "High Speed Rotor Controller", if the X-ray Tube
is rotating when touching the "OFF" push-button, the Generator will stop the
anode immediately and then the unit will be turned off (approx. 3 seconds). The
equipment only turns off if "Preparation" is not activated.
The proper way for switching OFF the Generator and Console
must be by touching on the "OFF" button on the Touch Screen,
never use the "ON / OFF Switch" of the Console.
IN THE EVENT OF AN EMERGENCY FORCIBLY DEPRESS
THE X-RAY ROOM "EMERGENCY OFF SWITCH" (USUALLY
A RED MUSHROOM-SHAPED SWITCH).
THIS SWITCH SHOULD BE LOCATED ON OR NEAR THE
X-RAY ROOM ELECTRICAL CABINET, USUALLY PLACED
NEAR THE GENERATOR CONTROL CONSOLE. MORE THAN
ONE OF THESE SWITCHES MAY BE PLACED AROUND THE
ROOM FOR GREATER ACCESSIBILITY.
TO ISOLATE THE EQUIPMENT FROM MAINS, TURN OFF THE
SWITCH LOCATED AT THE ROOM ELECTRICAL CABINET.
This button is the access to the "Service Mode" menus.
Only service personnel specifically trained on this medical X-ray equipment
should access to Service Mode for service tasks or maintenance of the
equipment.
0232F EN 20191213