DX-D 400
User Manual
Table 6-1 (cont.)
Error Codes
ERROR
DESCRIPTION
E41 to E46
System failure related to Dosimeter.
Collimator Error. A failure on the Automatic Collimator has
been detected (blades are full open or in movement during
E48
exposure, etc.)
E50
Exposure has been aborted by the Operator.
E51 to E93
System failure related to High Speed Rotor Controller.
E95
Exposure aborted by the AEC Rapid Termination.
X-ray order not acknowledged.
E97
The Receptor (Digital Detector) has not received or
acknowledged the X-ray order sent by the Generator.
E98
Service Mode Active.
138
WHAT TO DO
Press the "AEC Reset" push-button.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
If the equipment remains inoperative turn it OFF and call Field Service
Press the "AEC Reset" push-button.
Select the correct Ion Chamber or modify parameters.
Repeat the exposure.
If the equipment remains inoperative, turn it OFF and call Field Service.
Press the "AEC Reset" push-button.
Repeat the exposure.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
Press the "AEC Reset" push-button and call Field Service. This error
does not inhibit normal operation.
0232F EN 20191213