Note
0232F EN 20191213
ON / OFF / EMERGENCY OFF SWITCH: The Table and the Floor Mounted
Tube Stand are ON whenever the Generator Console is ON and the Table
Emergency OFF Switch is not depressed. The Table and the Floor Mounted
Tube Stand are OFF when the Generator Console is OFF or when the
Emergency OFF Switch is pressed.
IN THE EVENT OF AN EMERGENCY, TURN OFF THE TABLE
PRESSING FORCIBLY THE "EMERGENCY OFF SWITCH"
(RED MUSHROOM-SHAPED SWITCH) ON THE TABLE OR AT
THE ROOM ELECTRICAL CABINET.
Whenever the Table is powered on it takes twenty seconds
(start-up routine) previous to perform any motion.
CIRCUIT BREAKER: The Table includes a protective Circuit Breaker located
at the lower back cover. This breaker should remain in ON position to allow
Table operation.
TABLE-TOP LOCK SWITCH: This security switch disables Table-Top
horizontal movement avoiding any unexpected Table-Top motion (e.g. patient
steps on Pedal while sitting on Table before positioning or getting off the Table).
BEFORE PATIENT SITS OR GETS OFF THE TABLE PRESS
TABLE-TOP LOCK SWITCH TO AVOID PATIENT INJURY OR
DAMAGE TO EQUIPMENT.
THIS RAD TABLE PROVIDES A SAFETY SYSTEM UNDER
TABLE-TOP WHICH STOPS DOWN PEDAL MOVEMENT
WHEN TRAVEL FINDS AN OBSTACLE.