SECTION 4
4.1
FLOOR MOUNTED TUBE STAND
4.1.1
COLUMN ROTATION CONTROL
0232F EN 20191213
OPERATION
USE THE EQUIPMENT HAND-GRIPS TO CONTROL AND
DRIVE THE UNIT MOVEMENTS, NEVER PUSH DIRECTLY ON
THE RECEPTOR, X-RAY TUBE OR COLLIMATOR.
MONITOR THE SYSTEM MOVEMENTS WITH SPECIAL CARE.
AVOID ANY IMPACT OF THE SYSTEM ON FLOOR, CEILING
OR OTHER ELEMENTS IN THE ROOM.
SERIOUS DAMAGE TO THE EQUIPMENT.
MONITOR WITH SPECIAL CARE THE PATIENT POSITION
(HANDS, FEET, FINGERS, ETC.) AND USE THE PATIENT
HAND-GRIPS TO AVOID INJURY TO PATIENT CAUSED BY
TABLE-TOP MOVEMENTS. PATIENT HANDS MUST BE KEPT
AWAY FROM MOBILE COMPONENTS OF THE UNIT.
OPERATOR SHOULD NEVER MOVE TABLE-TOP FROM
LATERAL SIDE (PATIENT HEAD OR FEET). IT MAY CAUSE
FINGERPINCH UNDER THE TABLE-TOP.
INTRAVENOUS TUBING, CATHETERS AND OTHER PATIENT
CONNECTED LINES SHOULD BE ROUTED AWAY FROM
MOVING EQUIPMENT.
Operation controls for the Floor Mounted Tube Stand are located at the Control
Panel and on the Semicircular Base of the Column.
The Column rotation must be manually positioned by the Operator. Step on the
Pedal located at the Base of the Column and then rotate the Column to the
desired angle. Release the Pedal to lock the Column in position.
This motion can reach ±180
Rotation may be limited by cables. Avoid strain on the cables
when rotating.
o
o
with detents at 0
, 90
DX-D 400
User Manual
IT MAY CAUSE
o
o
and --90
.