GUIDANCE AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC IMMUNITY
The DX- -D400 radiographic room is intended for use in the electromagnetic environment specified below.
The customer or Operator of The DX- -D400 radiographic room should assure that it is used in such an environment.
Immunity Test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions
and voltage variations on power
supply input lines.
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
0232F EN 20191213
(IEC 61601-1-2:2014)
IEC 60601-1-2:2014
Test Level
8 kV contact
¦
2 kV,
 kV,
8 kV,
¦
¦
¦
15 kV air
¦
2 kV for power supply lines
¦
1 kV for input/output lines
¦
(100 kHz repetition frequency)
0.5 kV,
1 kV
¦
¦
line(s) to line(s)
0.5 kV,
1 kV,
2 kV
¦
¦
¦
line(s) to earth
0% U
for 0.5 cycle
T
o
o
o
o
o
at 0
, 45
, 90
,135
, 180
,
o
o
o
225
, 270
and 315
0 % U
for 1 cycle
T
o
at 0
70 % U
for 25/30 cycles
T
o
at 0
0% U
250/300 cycles
T
30 A/m
NOTE - U
is the a.c. mains voltage prior to application of the test level.
T
Compliance Level
8 kV contact
¦
2 kV,
 kV,
8 kV,
¦
¦
¦
15 kV air
¦
2 kV for power supply lines
¦
1 kV for input/output lines
¦
(100 kHz repetition frequency)
0.5 kV,
1 kV
¦
¦
line(s) to line(s)
0.5 kV,
1 kV,
2 kV
¦
¦
¦
line(s) to earth
0% U
for 0.5 cycle
T
o
o
o
o
o
at 0
, 45
, 90
,135
, 180
,
o
o
o
225
, 270
and 315
0 % U
for 1 cycle
T
o
at 0
70 % U
for 25/30 cycles
T
o
at 0
0% U
250/300 cycles
T
30 A/m
DX-D 400
User Manual
Electromagnetic
environment - guidance
Floors should be wood, concrete or
ceramic tile.
If floors are covered with synthetic
material, the relative humidity
should be at least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the This
X-ray System requires continued
operation during power mains
interruptions, it is recommended
that this X-ray System is powered
from an Uninterruptible Power
Supply or a battery.
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment.
35