DX-D 400
User Manual
4
X-Ray Generator:
Constant potential high frequency.
G
Three point control by selecting kVp, mA and ms, two point control
G
by selecting kVp and mAs (no AEC), or one point control by
selecting kVp with AEC operations.
Anatomical Programmer (APR) for six patient sizes (three adults
G
and three children), with pre-programmed anatomical views for
automatic selection. The operator may manually modify all the
original APR techniques and store them for later use.
Automatic Exposure Control (AEC).
G
Rapid Termination function.
G
Fail-safe function for the detection of the correct alignment of
G
the Tube-Collimator / Receptor with respect to the selected
workstation.
Two Buckys or DR Detectors can be directly connected to the
G
Generator.
Self-diagnosis indicators identify malfunctions in the system.
G
Tube protection circuitry prolongs Tube life and increases system
G
performance.
Equipped with closed loop control of X-ray Tube current, kVp and
G
filaments, which minimizes potential errors and the need for
readjustments.
Automatic line compensation due to closed loop operation of X-ray
G
Tube current and kVp.