ERROR CODES
0232F EN 20191213
TECHNIQUE ERROR: If activates during exposure it means that:
The exposure has been interrupted by "Security Timer"
because of a failure in the system. Call Field Service.
This error can also be shown:
after an APR technique selection to advise that exposure parameters
displayed on the Console are not the values stored for this APR
technique. Exposure parameters are adapted by the Generator to
another enable values.
if a failure on the Automatic Collimator has been detected (blades are full
open or in movement during exposure, etc.). In this case the indicator
lights continuously.
Error codes indicate the potential cause of a system failure. Error codes are
shown on the kVp Display at the same time an audio signal is emitted. Correct
the error cause and keep pressed the "AEC Reset" push-button till the Console
indication disappears. (Refer to Table 6-1).
All these error codes are preceded by the letter "E" (i.e., E01) and they will
enable the operator to indirectly convey the possible source of error to service
personnel. This may prevent the need for a service call or enable service
personnel to anticipate corrective actions prior to arriving on site.
DX-D 400
User Manual
135