STORE: All parameters of the APR techniques may be manually rewritten as
required by the operator. If the operator determines that some factors in an APR
technique should be re-programmed, perform the following procedure:
1.
Select an APR technique and modify the factors and selections of
Workstations or AEC which require to be re-programmed.
2.
Verify that all factors of the technique are at the required values.
3.
Touch the "Store" button to store the new technique. The selected
technique is now stored in the "APR Working File" and can be recalled
for future examinations.
This procedure only changes the technique values of the selected
Patient Size, if necessary repeat the procedure for other Patient
Sizes.
If an APR technique is to be stored with AEC parameters, a
suitable back-up Time (or mAs) MUST be stored by the operator
for this APR technique.
Since "Store" modifies the data stored in the "APR Working File", (according to
the configuration of the Unit), the Console requires the Operator Password to
allow the storage of the new data in the "APR Working File".
The default Operator Password is "1111", but it can be replaced by a new one
with the "Change Password" option in the "Edit" menu (refer to "Edit").
Enter the Operator Password and touch "OK" , then confirm or cancel the
operation.