AGFA DX-D 400 User Manual page 111

Hide thumbs Also See for DX-D 400:
Table of Contents

Advertisement

DX-D 400
User Manual
.
If after touching any of these buttons the technique value is
Note
blocked, it could mean that:
Radiographic Parameters Blockage. When any of the
maximum or minimum radiographic parameter limits are reached,
its related Display flashes accompanied by an information
message.
Generator Power Limit. If the Generator power limit (kVp x mA)
is reached, the kVp and mA Displays flash accompanied of an
information message.
Except in "3-Points" Working Mode, if the Generator power limit is
reached by increasing the mA up to a maximum possible value,
kVp could be increased up to its maximum value while mA value
may automatically decrease, as long as mAs value is kept
constant.
Space Charge. If a variation of the kVp or mA values means that
the space charge limit will be reached in the selected Tube, the
parameter is blocked, the kVp and mA Displays flash
accompanied by an information message.
Maximum Energy (60 kJ). Only in AEC mode, if a variation of the
parameters means that the maximum energy (60 kJ) will be
exceeded, the parameter is blocked, the kVp and mAs Displays
flash accompanied by an information message.
Instantaneous Power. If a technique reaches the instantaneous
power limit of the X-ray Tube (ratings limit or the X-ray Tube is
momentarily overheated), some techniques cannot be selected.
The kVp, mA and Time Displays flash accompanied by an
information message.
0232F EN 20191213
101