AGFA DX-D 400 User Manual page 66

Hide thumbs Also See for DX-D 400:
Table of Contents

Advertisement

DX-D 400
User Manual
.
Note
56
HORIZONTAL MOVEMENT: Activate this button to allow horizontal movement
of Column in order to achieve SID (Source-Image Distance) with respect to RAD
Wall Stand Receptor. Also this control is used to horizontally position Column
with respect to RAD Table Receptor. Release button to lock in position.
This movement is provided with detents related to the distance from the RAD
Wall Stand Receptor (that can be modified during the configuration of the
equipment). The detent point is shown at SID Display when the distance is
reached with respect to the RAD Wall Stand Receptor. The Display only shows
configured positions (a.e 110-150-180). Press this button again to continue
movement.
VERTICAL MOVEMENT: Activate this button to allow vertical movement of the
Tube-Collimator Assembly in order to achieve the desired SID with respect to
RAD Table Receptor. Also this control is used to vertically position
Tube-Collimator Assembly with respect to the RAD Wall Stand Receptor (in this
case the Display is not showing height). Release button to lock in position.
This movement is provided with a detent at 100 cm (40") (shown at SID Display)
when the distance is adjusted with respect to the RAD Table Receptor. Press
this button again to continue movement.
Horizontal and vertical movements always correspond to the
button arrow direction, that is, the buttons for horizontal and
vertical
movement
Tube-Collimator is turned 61
ROTATIONAL MOVEMENT: Activate this button to allow vertical rotation of
Tube-Collimator Assembly to the desired angle. Release button to lock in
position.
This movement has detents at 0
this button again to continue with movement.
interchange
its
function
o
left or right.
o
o
o
, 90
and --90
shown at Angle Display. Press
when
the
0232F EN 20191213