AGFA DX-D 400 User Manual page 57

Hide thumbs Also See for DX-D 400:
Table of Contents

Advertisement

SECTION 3
3.1
START UP WITH X-RAY GENERATOR CONTROL
0232F EN 20191213
START UP AND SHUTDOWN
The System should be powered by the same Room Electrical Cabinet where
the X-ray Generator is connected, that is, the whole System will be powered
from the same Electrical Cabinet.
To turn the System ON:
1.
Turn ON Room Electrical Cabinet Switch. The Emergency OFF Switch
must not be activated.
IN THE EVENT OF AN EMERGENCY FORCIBLY DEPRESS
THE X-RAY ROOM "EMERGENCY OFF SWITCH" (USUALLY
A RED MUSHROOM-SHAPED SWITCH).
THIS SWITCH SHOULD BE LOCATED ON OR NEAR THE
X-RAY ROOM ELECTRICAL CABINET, USUALLY PLACED
NEAR THE GENERATOR CONTROL CONSOLE. MORE THAN
ONE OF THESE SWITCHES MAY BE PLACED AROUND THE
ROOM FOR GREATER ACCESSIBILITY.
TO ISOLATE THE EQUIPMENT FROM MAINS, TURN OFF THE
SWITCH LOCATED AT THE ROOM ELECTRICAL CABINET.
2.
Check that the Emergency OFF Switch of the Table is not depressed and
the protective Circuit Breaker located at the lower back of the Table, right
side of the lower cover is in ON position to allow Table operation.
DX-D 400
User Manual
47

Advertisement

Table of Contents
loading

This manual is also suitable for:

5420/100

Table of Contents