DX-D 400
User Manual
Table 5-1 (cont.)
Error Codes
ERROR
DESCRIPTION
E19, E20
System failure.
E21, E22
Incorrect selection of the X-ray Tube.
E23
System failure.
E24
Bucky not ready for an exposure.
E27
System failure.
E28
AEC Reference set failure.
E32
Removable Receptor has not been replaced or inserted
E33
Serial Communication error.
Technique error. If it activates during exposure it means
that the exposure has been interrupted by the "Security
Timer" because of a system failure. Call Field Service.
This error can also be shown: after an APR technique
E34
selection to advise that exposure parameters displayed on
the console are not the values stored for this APR
technique. Exposure parameters are adapted by the
Generator to another enable values.
Door Open error. The X-ray room door is open when the
E35
X-ray equipment is in use.
Heat Units error. The X-ray Tube thermostat / pressurestat
is open due to the Tube housing being overheated (housing
E36
is too hot, wait for the housing to cool) or a thermostat /
pressurestat mal-function (housing is cool). Heat units may
raise to any value.
Tube Overload error. The technique selected is beyond the
X-ray tube ratings or present conditions of the X-ray tube
inhibit the exposure (anode overheated). Parameters for
next exposure may be temporally limited by the Generator
E37
(change the exposure values or wait for the X-ray tube to
cool). Check that heat units available are lower than the
calculated for the next exposure (heat units close to zero).
Reduce exposure factors or wait for the X-ray tube to cool.
E38
Wrong X-ray Tube orientation for Table Workstation
E39
Wrong X-ray Tube orientation for Wall Stand Workstation
E40
Alignment hardware error.
120
WHAT TO DO
Turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
Touch the "Reset" button.
Touch the "Reset" button
If the error code persists, turn the Generator OFF and ON.
p
,
If the equipment remains inoperative, turn it OFF and call Field Service.
Touch the "Reset" button.
Replace or insert the Receptor.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
In this case, the Generator can still work in "Direct" mode.
Touch the "Reset" button.
Check that communication cable between Generator and console is
properly connected.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
It is not necessary to touch the "Reset" button, the error indication will
It is not necessary to touch the "Reset" button the error indication will
disappear automatically.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
Touch the "Reset" button.
Correct the X-ray Tube orientation respect to the selected Workstation.
If the error code persists, turn the Generator OFF and ON.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
In this case, the Generator can still work in "Direct" mode and with the
other Workstation.
Touch the "Reset" button.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
In this case, the Generator can still work in "Direct" mode.
O
S
0232F EN 20191213