AGFA DX-D 400 User Manual page 134

Hide thumbs Also See for DX-D 400:
Table of Contents

Advertisement

DX-D 400
User Manual
6.1
RADIOGRAPHY AND GENERAL CONTROLS
6.1.1
POWER ON / OFF
6.1.2
WORKSTATIONS SELECTION
124
ON: The Generator is turned ON by pressing this push-button. This starts a
power-up routine which is shown on the Console (a.e. P05.3 = Vers.5, Rev.3).
After the power-up routine the last workstation used will be automatically
selected.
OFF: The Generator is turned OFF by pressing this push-button.
For Generators equipped with "High Speed Rotor Controller", if the X-ray Tube
is rotating when pressing the "OFF" push-button, the Generator will stop the
anode immediately and then the unit will be turned off (approx. 3 seconds). The
equipment only turns off if "Preparation" is not activated.
IN THE EVENT OF AN EMERGENCY FORCIBLY DEPRESS
THE X-RAY ROOM "EMERGENCY OFF SWITCH" (USUALLY
A RED MUSHROOM-SHAPED SWITCH).
THIS SWITCH SHOULD BE LOCATED ON OR NEAR THE
X-RAY ROOM ELECTRICAL CABINET, USUALLY PLACED
NEAR THE GENERATOR CONTROL CONSOLE. MORE THAN
ONE OF THESE SWITCHES MAY BE PLACED AROUND THE
ROOM FOR GREATER ACCESSIBILITY.
TO ISOLATE THE EQUIPMENT FROM MAINS, TURN OFF THE
SWITCH LOCATED AT THE ROOM ELECTRICAL CABINET.
The workstations are configured according to the customer preferences during
the installation procedure. The following push-buttons select their respective
workstation (illuminated push-buttons).
If a push-button is not configured during the installation, the respective
push-button(s) can not be selected during operation.
0232F EN 20191213