5.1.5
RADIOGRAPHIC PARAMETERS
.
Note
0232F EN 20191213
For Radiographic Operating Modes refer to Section 7.3.
RADIOGRAPHIC DISPLAYS: They are divided in the kVp, mA, mAs and
Time (s) Displays where the following data are shown.
kVp DISPLAY shows the radiographic kVp value selected for the technique.
mA DISPLAY shows the radiographic mA value selected for the technique.
mAs DISPLAY can show:
The radiographic mAs value selected for the technique.
When an exposure is made with AEC, it shows the actual mAs at the end
of the exposure, whenever the "Prep" button has not been released.
Time DISPLAY can show:
The Time value (in seconds) selected for the technique.
When an exposure is made with AEC, it shows the back-up Time during
the exposure and the actual Time at the end of the exposure whenever
the "Prep" button has not been released.
RAD Displays can also show:
the values (blinking) of the actual Time, the calculated mAs, and the
selected kVp and mA radiographic parameters of the last exposure, after
touching the "Reset Error" button.
If an exposure is aborted by releasing the exposure control during the
exposure, the Display shows the actual mAs and Time values, the
messages "Last Exposure Parameters" and "Error 50: Interrupted
Exposure", until the "Reset Error" button is touched to reset the error
condition.
DX-D 400
User Manual
99