Note
.
Note
Illustration 6-2
APR Storage Push-buttons
Example : press push-buttons: 1
0232F EN 20191213
APR TECHNIQUE CHANGES
The APR techniques are factory pre-programmed to standard technique sets.
All the parameters and selections for the APR techniques may be modified by
the operator and stored in the non-volatile memory for later use.
If the operator determines that some factors in an APR technique should be
re-programmed, use the following procedure:
1.
Select an APR technique and modify the factors and selections of
Workstations or AEC which require to be re-programmed.
2.
Verify that all factors of the technique are at the required values.
3.
Store the new technique by first pressing the selected Patient Size and
without releasing it press the Projection selected. (Refer to
Illustration 6-2.)
The newly selected technique is now stored in memory and can be
recalled for future examinations.
This procedure only changes the technique values of the selected
patient size, repeat the procedure for other patient sizes.
If an APR technique is to be stored with AEC parameters, a
suitable back-up Time (or mAs) MUST be stored by the operator
for this APR technique.
AP/PA
SKULL
LATERAL
SKULL
SEMIAX.
LATERAL
AXIAL
st
Medium Patient Size and Lateral Projection, to store the APR technique selected.
SPECIAL
DX-D 400
User Manual
131