AGFA DX-D 400 User Manual page 157

Hide thumbs Also See for DX-D 400:
Table of Contents

Advertisement

SECTION 9
9.1
ENVIRONMENTAL REQUIREMENTS
ATMOSPHERIC PRESSURE (hPa)
MIN
700 hPa
500 hPa
.
Note
9.2
X-RAY SYSTEM POSITIONERS
9.2.1
POWER LINE REQUIREMENTS
EQUIPMENT
RAD TABLE - - ELEVATING TABLE
RAD TABLE - - FIXED HEIGHT
TABLE
RAD WALL STAND
* NOTE: Power Line for the RAD Wall Stand is supplied by the X-ray Generator (110 -- 230 VAC, Single Phase, 50 / 60 Hz)
0232F EN 20191213
TECHNICAL SPECIFICATIONS
RELATIVE HUMIDITY (%)
MAX
MIN
1060 hPa
30%
TRANSPORT & STORAGE
1060 hPa
10%
These environmental conditions do not include the Digital
Detector. Refer to the Digital Detector Documentation.
VOLTAGE
110 - 230 VAC,
Single Phase
110 - 230 VAC,
Single Phase
* See Note
MAX
WORKING
75%
90%
MINIMUM INPUT
FREQUENCY
POWER
REQUIRED
50/60 Hz
0.6 kVA
50/60 Hz
0.4 kVA
0.5 kVA
DX-D 400
User Manual
AMBIENT TEMPERATURE
MIN
MAX
10 o C (50 o F)
35 o C (95 o F)
-20 o C (--4 o F)
70 o C (158 o F)
POWER
BRAKE POWER
OUTPUT
SUPPLY
0.18 kW
0.18 kW
0.10 kW (4A)
24 VDC, 2A
--
--
147