QUANTITATIVE INFORMATION
.
Note
2.9.1
FUNCTIONAL TESTS PERFORMED TO OBTAIN THE QUANTITATIVE INFORMATION
38
The following tables show the Quantitative Information associated
to this equipment according to the StandardIEC 60601- -1- -3:2008
and IEC 60601- -1- -3:2008+AMD1:2013. These tables illustrate
loading factors for image performance and supply Dose indication
examples. Therefore, they are an example of the adjustment of
Loading Factors, Focal Spot Selection, SID and Collimator
opening, which affect to the radiation quality or to the radiation
dose rate applied in normal use.
Equipment:
Rad Positioner with Ralco Collimator.
Instrumentation used:
Dosimeter: Vacudap
Dosimeter: Unfors
Rectangular Phantom made of Polymethyl-methacrylate (PMMA)
layers: 25 cm x 25 cm x 20 cm.
Test Details:
Minimum SID distance: 100 cm.
Maximum SID distance: 180 cm.
Open Collimator size: 13 cm x 13 cm (min.), 43 cm x 43 cm (max.)
The measurements were made with the exposure parameters shown on
the results table:
KVp Range: 40 KVp, 60 KVp, 80 KVp, 100 KVp, 125 KVp
mAs Range: 1 mAs, 2 mAs, 10 mAs, 50 mAs, 100 mAs
Performed measurements of Air Kerma or Air Kerma Rate at the
following designated positions:
-- Distance SID doses
-- Patient (Phantom) Entrance doses and Entrance doses Rate
-- Patient (Phantom) Output doses and Output doses Rate
-- Collimator Output doses