7. Load the Solutions cartridge into the Solutions station until it snaps into place and is level on the
deck.
Solutions station (room temperature)
1

Load the Recovery Tubes and Enrichment Cartridge v2

1. Load six Recovery Tubes (v2) into each Recovery centrifuge.
Recovery centrifuges
1
Before sealing each centrifuge, confirm that:
• The centrifuge is load-balanced with all required consumables.
IMPORTANT! The centrifuge must be load-balanced.
• The buckets are securely seated in the centrifuge rotors.
• The buckets are oriented correctly in the centrifuge so that they pivot outwards.
2. Place a Recovery Station Disposable Lid v2 over each centrifuge by lining up the tab with the
depression on the deck, then snap into place. Ensure that the lids snap completely into place by
applying firm downward pressure along the lid perimeter.
Ion 520
& Ion 530
ExT Kit − Chef User Guide