Initialization—General errors
Observation
Bottle leak check fails
64
S5 Systems setup and operation
Possible cause
• Bottle seal was not tight.
• Bottle was damaged or
defective.
Recommended action
1. Finger-tighten the bottles.
2. If the bottle continues to leak, replace the
bottle.
3. If leak check continues to fail, contact
Technical Support.