Section 7
Initial Operation
Cartridge Pack
Installation

Initial Rinse

7-2
Nanopure Life Science (UV/UF)
1. Open front (left side) door. Turn the cartridge pack so the caps are
facing you. Remove the caps.
2. Verify that each of the four posts on the cartridge pack has an o-ring.
3. Lift unit manifold upwards, insert cartridge pack and align the pack so
that it mates with the unit manifold.
4. Lower unit manifold until it is flush with the top of the cartridge pack.
5. Hand-tighten winghead screws securely.
6. Close door.
Figure 7-1. Cartridge Pack Installation
1. Attach the feed water line to the unit by snapping the quick disconnect
coupling into the quick disconnect coupling in the lower left back of
the unit.
2. Attach an atmospheric drain line (1/4" [.64 cm]) tubing by pushing
into the quick connect fitting on the lower left back of unit. See layout
connection under the "Water Connections" section.
3. Check to ensure the dispense knob is in the OFF (middle) position.
Manifold