Chapter 7 Clean the Ion Chef
7
Clean the Ion Chef
Instrument
4. Remove, then discard the box of used pipette tips from the waste tip position.
IMPORTANT! Handle the disposable reservoir in the waste tip position with care. During the run,
liquid waste collects in the reservoir. Dispose of the liquid waste by tipping the reservoir on one
corner and pouring the waste into an appropriate waste container:
IMPORTANT! Do not reuse the waste pipette tip rack. Always move the empty Ion S5
Cartridge from the new tip position to the waste tip position.
5. Move the empty Ion S5
6. Remove, then discard the
• Ion S5
• Ion S5
• Enrichment Cartridge v2
7. Remove, then discard the consumables from the Recovery centrifuges, including the:
• Recovery Station Disposable Lid v2
• Recovery Tubes v2
8. Remove the Chip Balance from the Chip-loading centrifuge. Do not discard.
9. Close the Chip-loading centrifuge lid.