Chapter 4 Start the Ion Chef
4
Start the Ion Chef
run
5. Follow the on-screen instructions. When prompted, close the instrument door by first lifting it
slightly to disengage the locking mechanism, then push down on the door until the locks engage.
IMPORTANT! Do not close the door by pulling it straight down from the open position. You must
lift the door slightly before you can close it. Ensure that both sides of the door are locked after
closing it.
Lift door first
1
Lower
2
After the door closes, the instrument vision system activates.
6. When prompted, tap Start check to begin Deck Scan. Wait while the instrument scans the
barcodes of all consumables and reagents to confirm their presence and compatibility.
During Deck Scan, the touchscreen displays warnings if the Ion Chef
or incompatible consumables. You must address all warnings before the run can start. After you
address each condition, tap Yes to continue.
IMPORTANT! The Deck Scan function is not a substitute for manual inspection of the reagents
and consumables on the Ion Chef
instrument operation, ensure that all consumables are installed correctly before you continue.
7. When the Deck Scan is complete, tap Next to display the Data Destination screen.
38
run
Instrument before starting a run. To ensure proper and safe
Press down to lock
3
Instrument detects missing
Ion 520
& Ion 530
ExT Kit − Chef User Guide