Chapter 5 Initialize the sequencer
5
Initialize the sequencer
7. If necessary, install a new Ion S5
Note: The Ion S5
8. Close the door, then tap Next.
The instrument confirms that the consumables and chip are properly installed and that the Ion S5
Cleaning Solution contains sufficient reagent to perform the post-run clean. Follow all on-screen
recommendations to ensure proper installation of required consumables.
IMPORTANT! If the allowed number of post-run cleans has been met, the instrument prompts the
user to replace the Ion S5
Note: If a Checking Reagent: Failed warning appears, see "Bottle leak check fails"
9. When initialization is complete (~50 minutes), tap Home.
The instrument is now ready for a sequencing run.
For information on recycling and disposal of used components, see "Reagent consumables disposal"
on page
52.
48
™
Cleaning Solution bottle.
™
Cleaning Solution bottle contains sufficient reagent to complete 4 cleanings.
™
Cleaning Solution bottle.
Ion 520
™
& Ion 530
™
ExT Kit − Chef User Guide
™
on page
64.