(continued)
Step or field name
Sample Name (Required)
Monitoring Thresholds
Ion 520
& Ion 530
ExT Kit − Chef User Guide
Did you select a Barcode Set in the Kits step?
• No – Enter the sample name for each sample loaded onto each chip in
the Sample Name (required) column, then enter or scan the corresponding
Ion Chef
Library Sample Tube barcode and the chip barcode in the Sample
Tube Label and Chip Barcode columns respectively.
• Yes – Enter the sample name for each sample in the Sample Name (required)
column, then select the barcode that is associated with each sample from the
dropdown list in the Barcode column.
Did you select the Ion Reporter Account and the associated Ion Reporter
workflow?
• No – For each sample name listed, make the appropriate selections in each
column in the table.
• Yes – For each sample name listed, also make the appropriate selections in the
Ion Reporter Workflow, Relation, Gender, and IR Set ID columns.
Set thresholds for Bead Loading, Usable Sequence, and Key Signal. In the Torrent
Suite
Software Monitor4Runs in Progress screen, an alert is displayed if the
values for a run fall below the selected thresholds.
Chapter 3 Create a Planned Run
Planned Run workflow key fields
Description
3
23