b. Grasp the chip by its edges, carefully lift the chip out of the bucket, then set it aside on a
clean, static-free surface. Return the bucket to the Chip-loading centrifuge.
1
2
3
Chip Adapter
1
Ion Chip
2
Bucket
3
4. Close the instrument door by first lifting it slightly to disengage the locking mechanism, then push
down on the door until the locks engage.
IMPORTANT! Do not close the door by pulling it straight down from the open position. You must
lift the door slightly before you can close it. Ensure that both sides of the door are locked after
closing it.
Lift door first
1
Lower
2
Ion 520
& Ion 530
ExT Kit − Chef User Guide
4
5
Press, then remove adapter
4
Remove from bucket carefully
5
Press down to lock
3
Chapter 4 Start the Ion Chef
Unload the chip for sequencing
run