Ventgard Cap Installation; Initial Cartridge Installation - Thermo Scientific EASYpure RoDi Operating Manual

Ultrapure water purification system
Table of Contents

Advertisement

Initial Cartridge
Installation
Ventgard Cap
Installation
Thermo Scientific
Refer to Figure 5-7 for the procedure below.
1. Open cartridge access door in the rear of the unit by pushing the door
latch back.
2. Remove a new RO carbon prefilter cartridge from its plastic bag.
3. Wet the o-rings on both end caps with water.
4. Press the upper end cap into the RO carbon
prefilter position until it bottoms out. The two
flanges on the upper end cap should be able to
slide down on each side of the keyway wall.
5. Lower the cartridge and insert the lower end
cap into the lower socket until it is firmly
seated.
6. Repeat Steps 2 - 5 with the EASYpure
ULTRApure cartridge, placing it in position 2.
Next, install the High Purity/Low TOC
cartridge, placing it in position 3.
7. Close and latch cartridge access door. This
serves to verify the cartridges have been
properly seated.
Caution The EASYpure RODIʼs cartridges must be installed in the proper
position. The upper end cap is the one with the right angle turn and two
flanges. The lower end cap extends straight out from the cartridge.
The Ventgard cap is shipped as a complete unit; installation involves
simply removing the new Ventgard cap from its plastic storage bag and
placing it on the reservoir. Refer to the Maintenance section for further
information.
Section 5
Section title
Flanges
O-rings
O-rings
Figure 5-7. Installation
s
EASYpure RoDi
5-7