Load Samples; Run The Gel - Thermo Scientific invitrogen E-Gel G8121ST User Manual

Power snap electrophoresis system
Table of Contents

Advertisement

Load samples

1.
Load prepared samples. Keep all sample
volumes uniform.
2.
Load prepared DNA ladder.
Note: Total loading volume for marker lanes
in double comb E-Gel
3.
Load 1X E-Gel Sample Loading Buffer or
deionized water in all empty wells.
Note: For E-Gel
E-Gel
4.
Run gels within 1 minute after loading
samples.

Run the gel

1.
Press Set up run to start E-Gel
2.
Select the E-Gel
type.
Use the up/down arrows to navigate through the menu.
3.
(Optional) For recurring experiments, select the last
used protocol.
Gel Type
E-Gel
EX Agarose Gel, 1% and 2%
E-Gel
EX Agarose Gel, 4%
E-Gel
EX Double Comb Agarose Gels,
1% and 2%
E-Gel
Agarose Gel with SYBR
1%, 2%, and 4%
E-Gel
Double Comb SYBR
Agarose Gels, 1% and 2%
E-Gel
CloneWell
II Agarose Gel, 0.8%
E-Gel
SizeSelect
II Agarose Gel, 2%
E-Gel
NGS
Agarose Gel, 0.8%
Reverse protocol for:
E-Gel
CloneWell
II Agarose Gel
E-Gel
SizeSelect
II Agarose Gel
E-Gel
Power Snap Electrophoresis System User Guide
agarose gels is 10 μL.
EX agarose gels, use 0.1X
Sample Loading Buffer.
protocol selection.
protocol corresponding to your gel
Recommended program
E-Gel EX 1-2%
E-Gel EX 4%
E-Gel EX 1-2%
Safe,
E-Gel 0.8-2%
Safe
E-Gel Double Comb
CloneWell 0.8%
SizeSelect 2%
E-Gel 0.8-2%
Reverse E-Gel
Default run time
Maximum run time
10 min
15 min
5 min
26 min
13 min
12 min
8 min
26 min
2 min
15 min
20 min
8 min
40 min
18 min
28 min
25 min
32 min
3 min