Chapter 4 Start the Ion Chef
4
Load the Ion Chef
System
3. Close the hinged cover of the Recovery centrifuges. Confirm that the port of each disposable lid is
positioned toward the rear of the instrument.
1
Recovery Tubes installed
1
Recovery Station Disposable Lids installed
2
Recovery centrifuge cover closed
3
Port
4
IMPORTANT!
·
Do not obstruct or place any object on top of the Recovery centrifuge cover.
·
Use only the supplied materials, including buckets and disposables, to run the centrifuges at
the rated speeds. Do not remove or change the rotors. To ensure normal operation, inspect the
buckets before each use.
4. Load the Enrichment Cartridge v2, then press down on the cartridge to ensure that it is level with
the instrument deck.
IMPORTANT! Ensure that the Enrichment Cartridge v2 is loaded so that the lettering on the
cartridge is right‑side‑up.
Enrichment station
1
Enrichment Cartridge v2
2
32
run
2
4
1
2
3
4
5
6
7
8
9
10
11 12
A
B
C
D
E
F
G
H
3
4
A
B
C
D
E
F
G
H
Lettering
3
Ion 520
& Ion 530
ExT Kit − Chef User Guide