Inspect and clean the Recovery centrifuges and buckets

1. Inspect the Recovery centrifuges, then clean the components if excess liquid is present.
Is liquid present?
No
Yes
IMPORTANT! Wear powder-free, nitrile gloves when cleaning the Recovery centrifuge.
2. Remove the buckets from the Recovery centrifuges. Clean the inside and outside of each bucket
using a lint-free wipe, then place the buckets on a clean, dry surface while you clean the
centrifuge.
Bucket
1
3. Use lint-free wipes to remove all fluid from inside the centrifuge bowl.
Inside rim of the centrifuge
1
4. Use lint-free wipes treated with 70% isopropanol to clean the following parts.
• Inside rim of the centrifuge.
• Bottom of the centrifuge bowl.
• Outside and inside of the centrifuge buckets.
Ion 520
& Ion 530
ExT Kit − Chef User Guide
Proceed to "Start the cleaning"
Clean the centrifuge bowl and buckets as described below.
IMPORTANT! Clean the Recovery centrifuges occasionally, only when excess
liquid is noticeable in the bowl and/or buckets. You do not need to clean the
centrifuges after every run.
Chapter 7 Clean the Ion Chef
Clean the Ion Chef
Action
on page
58.
1
2
Lint-free wipe
2
1
2
Bottom of the centrifuge bowls
2
System
7
Instrument
57