D/A Output; Table 5-1: Test Channel; Table 5-2: Diagnostict - Teledyne 401 Instruction Manual

O3 calibrator
Table of Contents

Advertisement

Teledyne API Model 401 O
Table 5-1: Test Channel Output
Test #
Name
1
NONE
2
O3 DETECT
3
REF DETECT
4
SAMP PRESS
5
SAMP FLOW
6
SAMP TEMP
7
ALAMP TEMP
8
O3 FLOW
9
O3 LAMP TEMP
10
CHASSIS TEMP
11
DCPS VOLT
Table 5-2: Diagnostic Tests
Test #
Name
1

D/A OUTPUT

5.2.1 D/A Output
This test cycles 3 of the analog output channels from 0% to 100% of Full Scale in 20% FS steps.
It starts by outputting 0 volts to all four channels and displaying a 0% button. Then, every five
seconds, the output is increased 20% FS and the button is changed accordingly. Thus, the button
(and the analog outputs) will cycle through the following values:
0%, 20%, 40%, 60%, 80%, 100%, 0%
To pause the output at the current voltage, press the n% button. To resume automatic cycling,
press the n% button again.
Photometric Calibrator Instruction Manual, 01124, Rev. J1
3
Purpose
No output
Outputs O
detector
3
Outputs IZS ref. det.
Outputs sample press.
Outputs sample flow
Outputs sample temp.
Outputs ana lamp temp.
Outputs O
flow
3
Outputs O
lamp temp
3
Outputs chassis temp
Outputs DC power
Type
Analog
5-2
Purpose
Tests analog outputs