Troubleshooting and Maintenance
Prompt
Air In Line
Dose Out of Range
Entry Invalid
Instrument Self-check is Due
Please Eject the Set
Invalid Entry Rate Out of Range
Latch Open
Maintenance Reminder
New Baseline Set
Occlusion Downstream
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
Alarms, Alerts, Prompts (Continued)

Prompts

Air detector detected air prior to
starting infusion or is in poor
contact with set.
Calculated dose is outside
allowable range.
An invalid value was entered
during programming.
Instrument or channel did not
perform self-check within
programmed interval.
Instrument calculated a rate less
than
0.1
Latch is open—prior to starting an
infusion.
Periodic maintenance interval
elapsed.
Maintenance Reminder feature is
on.
A new Manual Pressure Baseline
was successfully set.
Manual Pressure Baseline feature
is on.
A very high pressure exists in fluid
line while baseline is being set.
Pressure Baseline feature is on.
Meaning
mL/h.
Alarms, Alerts, Prompts
Response
Press continue soft key to allow
infusion to continue. An alarm
occurs if air detector detects an air
bubble larger than configured
threshold. Verify that set is loaded
correctly. Prime and reload set or
remove air. Reshape tubing to
ensure optimum contact with
sensor.
Ensure air-in-line sensors are
thoroughly cleaned.
Verify and reenter settings.
Press
key or
CLEAR
0
entry. Enter appropriate value.
If set is loaded: Eject set, wait five
seconds, and then reload set.
If no set is loaded: Load set, wait
one minute, and then eject set.
Wait five seconds and then reload
set.
Verify and reenter settings.
Close latch fully to left.
Notify Biomedical Engineering
department. If desired, press
continue soft key to temporarily
bypass reminder.
Baseline remains set until a new
manual baseline is set, instrument
is turned off, or latch is opened.
Remove source of high pressure
and repeat setting of pressure
baseline.
key to clear
113