Rate Accuracy Qualification Test - CareFusion Alaris SE User Manual

Table of Contents

Advertisement

Check-In and Configuration
Getting Started
Check-In and Configuration (Continued)

Rate Accuracy Qualification Test

This procedure is to be used only when testing an instrument
during a new instrument check-in or when just received from
the service depot center. This test is to verify that damage or
changes to the instrument did not occur during shipment and
handling.
Rate accuracy should be tested using a Model
80VCS
Calibration Set. The system is designed to produce overall
accuracy of
for rates greater than
mL/h and up to
±5%
1
mL/h, and
for rates equal to or less than
mL/h,
999.9
±6.5%
1
of the time with
confidence (see Trumpet and Start-
95%
95%
Up Curves on page 100 for additional information). The
system performance with a calibration set produces a smaller
variability. In order to ensure overall accuracy is achieved, new
instruments are tested to an accuracy of
with the
±3%
Model
set during a new instrument check-in.
80VCS
Due to the dynamic monitoring feature, the rate is varied during
operation. For this reason, CareFusion does not recommend
using automatic testers to check rate accuracy. Generally,
these devices collect small samples and might cause the
results to be incorrect, even though the instrument is accurate.
Do not use the Model
Calibration Set for more than
80VCS
rate verification runs (
rate calibration number changes).
30
15
Keep track of the number of times the set is used by recording
each use on the
insert or on a separate record.
80VCS
4
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230