CareFusion Alaris SE User Manual page 120

Table of Contents

Advertisement

Troubleshooting and Maintenance
Prompt
Ok Entry
Panel Locked
Place on Hold to Change
Place on Hold to Set Pressure
Baseline
Press and Hold Key to Turn Off
Pressure Limit XXX mmHg
(XXX represents configured
maximum pressure)
Pressure Limit Must Be Less
Than or Equal to XXX mmHg
(XXX represents configured
maximum pressure)
Pressure Unstable Cannot Set
Baseline
114
Alarms, Alerts, Prompts (Continued)
Prompts (Continued)
Attempt was made to go to
another page before pressing ok
soft key.
A key was pressed.
Panel lock feature is on.
A key was pressed during
SET PRESSURE BASELINE
function was selected while
running.
Pressure Baseline feature is on.
POWER
An elevated pressure was present
in fluid path when pressure
baseline was established. This
might reduce maximum available
pressure range.
Attempt was made to increase
pressure alarm limit to a level
higher than configured maximum
pressure.
Excessive variation in pressure
due to motion, flow from other
instruments, or blood pressure
prevents accurate setting of
pressure baseline.
Pressure Baseline feature is on.
Meaning
KVO
key was pressed.
Alarms, Alerts, Prompts
Response
Verify that all parameters are
correct and press ok soft key.
Turn panel lock off to access panel
controls. Panel lock key is located
behind handle.
.
Channel must be on hold to make
changes.
Place instrument on hold before
performing manual
operation.
BASELINE
Press and hold
display turns off.
Reload administration set and verify
that no obstruction exists which
could cause excess pressure.
If Pressure Baseline feature is
on, repeat manual setting of
pressure baseline.
If Pressure Baseline feature is
off, restart infusion to
automatically set pressure
baseline.
Choose a pressure alarm limit that
is less than, or equal to, configured
maximum pressure.
Reduce or temporarily remove
sources of variation while
performing manual baseline setting
operation.
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
SET PRESSURE
key until
POWER

Hide quick links: