Power On And Off; Audio Adjust - CareFusion Alaris SE User Manual

Table of Contents

Advertisement

General Setup and Operation
1.
Press
key.
POWER
Instrument performs a self test.
All indicators and displays momentarily light.
An audio tone sounds.
Hold indicator flashes.
System startup page is briefly displayed.
example display represents current software revision;
ID No. is instrument serial number.
2.
Press and hold
This feature allows the audio volume to be adjusted for alarms,
alerts, and
tone to either High, Medium, or Low if all audio
KVO
volume levels are enabled in the hospital configuration settings.
Press
soft key next to lower
Audio volume level displays in lower
Low
Medium
High
Instrument can be configured to enable only Medium and
High, or High audio volume levels.
68
Start-Up (Continued)

Power On and Off

key until display turns off.
POWER

Audio Adjust

display.
LCD
in
2.XX
A B
OPTIONS
Model 7XXX
© 1998 Alaris Medical
SW Rev.: 2.XX
ID No.: XXXXXXXXX
display.
LCD
WARNING
Each time the instrument is turned
on verify and/or set the monitoring
mode, resistance alert, and/or
pressure alarm limit. If the
monitoring mode, resistance alert,
and/or pressure alarm limit are not
verified, the instrument might not be
operating with the desired occlusion
detection parameters.
CAUTION
Appearance of lines and/or
dots that remain on constantly
when the instrument is powered
on might indicate improper
functioning of the Main Display.
Although the instrument is
functioning, return it to qualified
service personnel.
for anomalies
Inspect
LCD
(improperly lit or unlit pixels).
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
Start-Up

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

7100713072007230

Table of Contents