Programming
The channel must be on hold or the last dose complete.
1.
Press menu soft key.
2.
To return to primary setup page, press Quit Program soft
key.
Primary setup page parameters might be different than
those for the Multi-Dose program. Verify all settings prior
to resuming an infusion.
This is an optional feature that is configured by the facility.
Options are enabled through the instrument configuration
settings.
The Multi-Step feature allows a sequential drug delivery
program (up to nine steps) to be set, delivering volumes of fluid
at different rates during each step. This allows the instrument
parameters to be set up once and to deliver a sequence,
eliminating the need to change the rate and
infusion step.
The infusion may be programmed in either rate and volume, or
volume and time. At completion of the last programmed step,
the channel switches to the preset
current rate, whichever is less.
1.
Select desired channel, as necessary. Channel must be
on hold in primary mode.
2.
Press
OPTIONS
3.
Press Multi-Step soft key.
4.
Press Enter New Program soft key.
40
Multi-Dose (Continued)

Quit Multi-Dose

Multi-Step

rate or remains at the
KVO

Program

key.
after each
VTBI
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
Multi-Step