Programming
Change Parameter During Infusion
Continuous infusion parameters (Rate or
changed without pausing the infusion, and
1.
Select applicable channel, as necessary.
2.
Press soft key next to parameter to be edited.
Current value is highlighted.
3.
Perform one of following steps:
•
To change rate or
•
To reset
to
VI
4.
To accept change, press
When the primary
VTBI
automatically switches to the configured
rate, or remains at the current infusion rate, whichever is less.
•
Programmed infusion rate continues to be displayed in
Main Display.
•
flashes in infusion status bar.
KVO
•
alert tone sounds (may be silenced for two minutes
KVO
using Silence key).
•
message flashes in Main Display.
VTBI=0
Resume Infusion from KVO Mode
1.
To place channel on hold, press
2.
Press
soft key.
VTBI
is highlighted.
VTBI
3.
Enter desired
VTBI
4.
Press
key.
ENTER
5.
To resume infusion, press
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
Primary Infusion (Continued)
VTBI
, enter a new value.
VTBI
mL, press
0.0
CLEAR
key.
ENTER
KVO Mode
reaches
mL, the instrument
0.0
KVO
/
RUN
.
/
RUN
HOLD
) may be
may be cleared.
VI
or
(zero) key.
0
(keep vein open)
key.
HOLD
key.
Primary Infusion
21