Table of Contents

Advertisement

Models 7100/7130 and 7200/7230, Rev2.X
RUN
POWER
HOLD
ml/hr
OPT
PRI
HLD
SEC
KVO
PRI
SEC
OPTIONS
1
2
3
4
5
6
8
9
7
.
0
Clear
Enter
User Manual
®
Alaris
SE Pump
A
RUN
RUN
POWER
POWER
HOLD
HOLD
ml/hr
ml/hr
KVO OPT
HLD PRI
SEC
PRI
SEC HLD
OPT KVO
PRI
SEC
A B
OPTIONS
1
2
3
5
4
6
7
8
9
0
Clear
Enter
April 2011
B

Advertisement

Table of Contents
loading

Summary of Contents for CareFusion Alaris SE

  • Page 1 User Manual ® Alaris SE Pump Models 7100/7130 and 7200/7230, Rev2.X April 2011 POWER HOLD POWER POWER HOLD HOLD ml/hr ml/hr ml/hr KVO OPT HLD PRI SEC HLD OPT KVO OPTIONS OPTIONS • Clear Enter Clear Enter...
  • Page 2 Technical Support and Customer Service - International (Maintenance and service information support. Product return, service assistance, and order placement) www.carefusion.com/customer-support/customer-service DIR 10000149477 © 2006-2011 CareFusion Corporation or one of its subsidiaries. All rights reserved. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 3: Table Of Contents

    Resume an Interrupted Multi-Dose ....................38 Quit Multi-Dose ..........................40 Multi-Step .............................. 40 Program ............................40 View Steps or Make Changes ....................... 44 Resume an Interrupted Infusion ....................46 Quit Multi-Step ..........................47 User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 4 Symbols and Terms ..........................97 Trumpet and Start-Up Curves ....................... 100 Flow Characteristics Under Varying Delivery Conditions .............. 101 Pressure Mode ..........................102 Resistance Mode ........................... 104 High Resistance Mode ........................106 User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 5 Prompts ............................113 Inspection Requirements ........................116 Cleaning ..............................116 Service Information ..........................117 Technical Support .......................... 117 Regulations and Standards Compliance ............................119 Electromagnetic Environment ......................119 Standards ............................126 Trademarks ............................126 User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 6 T H I S P A G E I N T E N T I O N A L L Y L E F T B L A N K User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 7: Getting Started Introduction

    ), subcutaneous, epidural, enteral, and irrigation of fluid spaces. The Alaris SE pump is available as either a single or a dual channel pump. The dual channel pump is a two-channel device intended to deliver multiple infusions to a single patient.
  • Page 8 Standards on page 119. Contraindications: None known. Warnings and Cautions: Warnings and cautions provide information needed to safely and effectively use the Alaris SE pump. See Warnings and Cautions on page 73. is an alert to an imminent hazard which DANGER could result in serious personal injury and/or product damage if proper procedures are not followed.
  • Page 9: Unpacking

    Maximum battery capacity, as well as gauge accuracy, is in the facility are responsible for reached after several charge/discharge/recharge cycles in unpacking the instrument and the refresh process. CareFusion recommends that the ensuring that the battery is fully battery be fully charged/discharged/recharged, using the charged before placing the refresh cycle, before placing the instrument in use.
  • Page 10: Rate Accuracy Qualification Test

    80VCS Due to the dynamic monitoring feature, the rate is varied during operation. For this reason, CareFusion does not recommend using automatic testers to check rate accuracy. Generally, these devices collect small samples and might cause the results to be incorrect, even though the instrument is accurate.
  • Page 11 Close AccuSlide flow regulator clamp. Connect output of set to one side of three-way stopcock. Load set into instrument and close latch. Verify no fluid flow or drops falling in drip chamber. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 12 Rate Accuracy Specification. The instrument might need to be run to prime the line to the zero level of the burette (step 14). User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 13 A = Vcollected ÷ Vexpected Step 1: Step 2: B = A x 100 Step 3: % Error (round % Error to nearest tenth of a percent) = B – 100 A passing result is 0.0±3% User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 14: Alternative Rate Accuracy Qualification Test

    Due to the dynamic monitoring feature, the rate is varied during operation. For this reason, CareFusion does not recommend using automatic testers to check rate accuracy. Generally, these devices collect small samples and might cause the results to be incorrect, even though the instrument is accurate.
  • Page 15 14. Press key to start primary infusion. Infuse until RUN/HOLD tubing is fully primed—approximately one minute. 15. Press key to stop infusion. RUN/HOLD The factory default for the Monitoring Options mode is Pressure. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 16 80VCS following conclusions is determined: • Instrument passes rate verification and calibration is not needed. • Rate calibration number was changed and instrument now passes verification. • Mechanism replacement is required. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 17: Set Sensor Check/Pressure Calibration Verification

    If result is outside a range of - to + from 5.5% 7.0% expected volume, return instrument to CareFusion for repair or replace mechanism. 27. Empty flask and reset scale to zero, in preparation for next test. Set Sensor Check/Pressure Calibration Verification...
  • Page 18: Functional Test

    VTBI Press key to begin infusion. RUN/HOLD Clamp off line just above instrument—about two inches—to simulate an upstream occlusion. Verify that instrument stops running, alarms, and displays within seconds. OCCLUSION UPSTREAM User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 19: Flow Stop Test

    With all tubing clamps open and fluid container two or more feet above instrument, verify that no fluid flows through set. Open latch and remove set. Verify that no fluid flows through set. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 20: Ground Current Leakage Test

    Close the latch(es) when the instrument is not in use. Administration Set Information General The Alaris SE pump uses a wide variety of AccuSlide flow regulator administration sets. The sets dedicated for use with the Alaris SE pump are designed for use with the instruments as well as for gravity-flow stand-alone use.
  • Page 21: Smartsite ® Needle-Free Valve

    For multiple syringes, swab prior to each syringe access. • Replace every hours or activations, whichever comes first. Exception: For infusions of blood, blood products, or lipid emulsions replace every hours. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 22: Prepare Solution Container And Set

    • Use only sets dedicated for use Slide AccuSlide flow regulator thumb clamp down until an with the Alaris SE pump. The audible click verifies that it is in fully closed position. use of any other set might cause improper instrument operation,...
  • Page 23: Load Set

    (up) position prior to starting infusion. free from kinks and installed Attach set to patient’s vascular access device. correctly in the instrument. Verify flow from container after starting infusion. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 24: Remove Set

    Send the instrument to qualified service personnel. Change Solution Container Place channel on hold. Remove empty solution container. Spike new container. ⅔ Ensure that drip chamber is filled to full. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 25: Programming

    0.0 mL Primary Limit=600 mmHG Perform one of following steps: • If current primary infusion rate is appropriate, press key. ENTER • Enter new infusion rate and press key. ENTER is highlighted. VTBI User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 26: Pause And Restart Infusion

    , pressing the key, and then pressing the VTBI CLEAR key. Or, the primary can be deleted from the primary VTBI ENTER mode setup page (see Configurable Options on page 92). User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 27: Change Parameter During Infusion

    Resume Infusion from KVO Mode To place channel on hold, press key. HOLD Press soft key. VTBI is highlighted. VTBI Enter desired VTBI Press key. ENTER To resume infusion, press key. HOLD User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 28: Clear Volume Infused

    • The secondary administration set must be primed prior to beginning the secondary infusion. • Verify that the secondary systems are properly set up, to ensure that there is a proper flow. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 29: Setup

    Using hanger(s) provided with secondary administration set, lower primary fluid container until bottom of secondary container is at least inches above fluid level in primary 9½ container. Hanger Secondary Container at least 9 " ½ Primary Container User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 30: Infusion

    The secondary mode must be set to in the instrument system configuration settings. If the secondary mode is , an invalid keypress tone sounds and "Secondary Not Allowed" is displayed when the key is pressed. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 31: Change Parameters During Infusion

    Enter a new value and press key. ENTER Change Primary Settings Select desired channel, as necessary. The channel does not need to be on hold to change the rate or VTBI values. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 32 , pressing the key, and then pressing the VTBI CLEAR key. Or, the primary can be deleted from the primary VTBI ENTER mode setup page (see Configurable Options on page 92). User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 33: Drug Specific Dose Rate Calculator (Drc)

    When a drug amount is greater than units (Un), a 10,000 is used to indicate a value multiplied by (for example, 1,000 1,000,000 1,000K • cannot be used in conjunction with secondary or other operating modes. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 34: Program

    DOSE RATE MENU Review/Resume Enter New Program Quit Program The channel must be infusing in the primary mode or on hold in the primary mode, secondary mode, or a loading dose program. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 35 Dose units cannot be changed. The weight and height unit selections are displayed only if they are appropriate for the selected drug. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 36 ○ Weight-based dosing: Press key to ENTER advance to time segment. ○ Non-weight-based dosing: Press soft key to clear segment, and then press key (weight ENTER field disappears). User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 37 Verify that all parameters are correct and press ok soft key. Enter a value and then press key. VTBI ENTER User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 38: Loading Dose

    Verify that the primary mode parameters are correct prior to accessing the Loading Dose option. Program Select desired channel, as necessary. Channel must be on hold in primary mode. Press key. OPTIONS User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 39 To reset volume infused to mL, press CLEAR (zero) key. To accept new value and begin infusion, press ENTER key. When loading dose infusion is complete, Loading Dose Complete is displayed and primary infusion resumes. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 40: Multi-Dose

    Press Enter New Program soft key. OPTIONS MULTI-DOSE MENU Review/Resume Enter New Program Quit Program Enter infusion rate and press key. ENTER OPTIONS Rate= ----- mL/h VTBI/Dose= ----- mL No. Doses= -- Frequency= -- h menu User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 41 If Dose Complete Alert Option is enabled, DOSE VTBI/Dose=100.0 mL page appears. No. Doses= 4 COMPLETE ALERT OPTION Frequency= 6 h menu 10. Press On or Off soft key, as applicable. OPTIONS DOSE COMPLETE ALERT OPTION menu User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 42 Enter number of hours until first dose and press ENTER key. OPTIONS Enter number of minutes ( ) until first dose and press key. ENTER Until First Dose VI= 0.0 mL menu start timer User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 43 VTBI=100.0 mL 100.0mL@ 250.0mL/h Q 6h x 4 VI=100.0 Adult ICU Change Time Interval Until Next Dose Press stop timer soft key. OPTIONS 12 min Until Dose 2 of 4 stop timer User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 44: Resume An Interrupted Multi-Dose

    Resume an Interrupted Multi-Dose Select desired channel, as necessary. Return To Multi-Dose? page is displayed. Press yes soft key. OPTIONS Pressing no soft key returns screen to primary setup page. Return To Multi-Dose? User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 45 To begin timer’s countdown to delivery of next dose, press start timer soft key. When final dose is complete: • Dose of Complete displays until user takes action. • No Continuous Infusion or KVO Rate occurs. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 46: Quit Multi-Dose

    Program Select desired channel, as necessary. Channel must be on hold in primary mode. Press key. OPTIONS Press Multi-Step soft key. Press Enter New Program soft key. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 47 To approve all displayed information and advance to OPTIONS of infusion sequence, press ok soft key. STEP 2 STEP 1 Rate= 25.0 mL/h VTBI= 5.0 mL Time= 0 h 12 min menu User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 48 VTBI= 130.0 mL Time= 1 h 42 min 25.0 menu To accept page, press ok soft key. STEP TOTALS OPTIONS STEP TOTALS (3) VTBI= 130.0 mL Time= 1 h 42 min 0.0 mL menu User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 49 To set up each additional step of infusion sequence, repeat steps through When all steps have been entered and accepted, press done soft key. Review page displays three sequence steps at a time. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 50: View Steps Or Make Changes

    OPTIONS Rate= 50.0 mL/h Press soft key. Remaining time and are displayed for a VTBI VTBI= 25.0 mL short interval. Totals Remaining VTBI= 125.0 mL Time= 1 h 30 min Adult ICU User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 51 Enter new value and press key. ENTER To accept changes, press ok soft key. To approve review page and page, press STEP TOTALS ok soft key. To resume infusion, press run soft key or key. HOLD User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 52: Resume An Interrupted Infusion

    If resuming an infusion on a dual channel instrument with an infusion currently running, the display goes directly to Return To Multi-Step? during startup. Pressing no soft key returns screen to primary setup page. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 53: Quit Multi-Step

    To return to primary setup page, press Quit Program soft key. Primary setup page parameters might be different than those for the Multi-Step program. Verify all settings prior to resuming an infusion. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 54 Multi-Step Programming T H I S P A G E I N T E N T I O N A L L Y L E F T B L A N K User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 55: Dynamic Monitoring System

    All features and options in this section are shown as enabled. WARNING Options are enabled through the instrument configuration • The Alaris SE pump is designed settings. to stop fluid flow under alarm conditions. Periodic patient The Dynamic Monitoring System provides the ability to monitor...
  • Page 56: Monitoring Options-General

    Resistance Trend Monitoring Options Dose Rate Calculator Set Pressure Baseline return page Precision Flow: In fixed and adjustable pressure modes, the Alaris SE pump provides enhanced flow continuity at rates below 50 mL/h. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 57 Resistance alert limit may be adjusted using the soft keys located below the arrow symbols (see Resistance Alert on page 54). High Resistance alert limit may be adjusted using the soft keys located below the arrow symbols. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 58: Monitoring Options-Resistance Mode

    OPTIONS is displayed and audible tone sounds. Rate= 125.0 mL/h VTBI= 1000.0 mL 0.0 mL Checking Line • If Auto Restart Plus feature is off, OCCLUSION OPTIONS alarm occurs. DOWNSTREAM OCCLUSION DOWNSTREAM User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 59: Auto Restart Plus Feature

    After the programmed number of restarts has occurred or the -second Checking Line period has been exceeded, the channel immediately alarms OCCLUSION when resistance or pressure conditions DOWNSTREAM indicate an occlusion. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 60: Resistance Alert

    Resistance Alert • Resistance alert level marker increases above current measured resistance value. • Resistance rises to , initiating a Checking Line or 100% condition. OCCLUSION DOWNSTREAM User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 61: Resistance Trend Graphs

    A tick mark ( ) at top of graph indicates an occlusion. • When viewing Resistance Trend Graphs in High OPTIONS Resistance mode, is displayed under graph. HI RESIST 100% -12h HI RESIST return clear time User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 62 The following events also turn off trend graph: • Pressing HOLD • An alarm • Dual channel instrument: ○ Pressing split screen key ( ○ Replaced with a split screen display after one minute if both channels are infusing User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 63: Monitoring Options-Pressure Mode

    The Adjustable Pressure mode allows control of the pressure alarm limit. If the condition is not cleared, the alarm OCCLUSION DOWNSTREAM occurs and infusion stops until manually restarted. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 64: Adjustable Pressure Alarm

    The first activation of for a new infusion HOLD automatically establishes a pressure baseline based on current system pressure. An optimal baseline is maintained upon subsequent activations of , as HOLD follows: User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 65 When multiple instruments are infusing through a common small diameter catheter, pressure measurement accuracy can be optimized by temporarily stopping all infusions, and then restarting all instruments beginning with the instrument delivering at the lowest rate. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 66: Pressure Baseline

    Select applicable channel, as necessary. Pressure bar graph is not shown when split screen display is active. Press key. HOLD All infusions connected to channel being base-lined must be on hold. Press key. OPTIONS User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 67 True baseline pressure will be zero or within a few mmHg of zero. If not, and the pressure is unstable, allow the pressure to drop to the lowest level and then repeat the Set Pressure Baseline process. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 68: Pressure Trend Graphs

    Setting a manual Pressure Baseline displays a horizontal real- time bar graph and numeric pressure readings. The vertical line on the pressure bar graph visually indicates the pressure alarm limit. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 69 To clear graphed data, press clear soft key. Press ok soft key. OPTIONS All trend data will be cleared return All data is cleared from graphs. OPTIONS -12h PRESSURE return clear time User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 70: Upstream Occlusions Detection

    (such as tubing) is free of obstructions, that any clamp is open, and that the blue thumb clamp on the AccuSlide flow regulator is in the open (up) position. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 71 Checking Line message for 10 seconds, to confirm or rule out presence of an occlusion. If occlusion condition is determined not to exist, flow resumes and no alarm is produced. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 72 Dynamic Monitoring System T H I S P A G E I N T E N T I O N A L L Y L E F T B L A N K User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 73: General Setup And Operation

    Change Pole Clamp Orientation Press and hold rotation lever. Lever Reposition clamp. Release lever at desired position. The illustrated pole clamp knob might not reflect the knob in use on the instrument. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 74: Power On And Off

    • Audio volume level displays in lower display. Medium High • Instrument can be configured to enable only Medium and High, or High audio volume levels. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 75: Tamper Resist-Panel Lock Feature

    0.0 mL on, channel select and split screen keys can be used to view settings. Panel Locked • To turn panel lock feature off, press and hold until disappears from lower display. GOLD User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 76: Respond To Maintenance Reminder

    Disconnecting the nurse call cable from the wall or turning off the instrument activates the nurse call system. Disconnecting the nurse call cable from the instrument will not activate the nurse call system. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 77: If An Alarm Occurs

    RS-232 connector. Connect to a Computer Press key. OPTIONS To advance to next page, press page soft key. Press Computer Link soft key. Press Monitor soft key. Press ok soft key. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 78: Disconnect From A Computer

    To advance to next page, press page soft key. Press Computer Link soft key. Press Off soft key. Press ok soft key. Disconnect cable from port on instrument RS-232 RS-232 rear panel. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 79: General Information Warnings And Cautions

    • Do not port the Alaris SE pump with a gravity flow infusion system. The use of positive displacement infusion devices ported together with gravity flow infusion systems into a common IV site can impede the flow of common gravity- only systems, affecting their performance.
  • Page 80 • Hyperbaric Chamber Operation: ○ The Alaris SE pump is not certified for use in oxygen- enriched environments. ○ The healthcare facility's hyperbaric safety director is responsible for all equipment used in the hyperbaric chamber environment.
  • Page 81: Administration Sets

    Discard if packaging is not intact or protector caps are unattached. • Use only sets dedicated for use with the Alaris SE pump. The use of any other set might cause improper instrument operation, resulting in an inaccurate fluid delivery or other potential hazard.
  • Page 82: Electromagnetic Compatibility

    Use only sets dedicated for use with the Alaris SE pump, without a ‘Y’ connector or injection port, for epidural infusions. ○ Epidural administration of anesthetic drugs: Use indwelling catheters specifically indicated for short-term hours or less) anesthetic epidural drug delivery.
  • Page 83: Battery Management System

    • Portable and Mobile RF communications can affect Medical Electrical Equipment. • The Alaris SE pump is intended for use under the supervision of healthcare professionals only. This is a Class device without the use of the Model CISPR 11...
  • Page 84: Battery Power Gauge And Indicator

    Use only • Do not open, incinerate, or CareFusion batteries. short-circuit the battery. • Worn-out batteries must be properly disposed of according to facility protocol. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 85: Flow Sensor

    In this event, unplug the flow ○ Upper surface of flow sensor sensor from the instrument. should be slightly below drop- forming orifice but above level of fluid in drip chamber. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 86 Routinely clean flow sensor with warm water while CAUTION actuating slider, and then dry thoroughly. Do not use solvents or cleaning agents. Damage to plastic parts of the flow sensor might occur. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 87: Features And Displays

    New alarm or alert reinstates Clear Enter audible tone. CLEAR Key: Clears selected numeric value. Audio Volume Key: Sets audio volume for alarms, alerts, and KVO tone. Numeric Keypad: Enters and changes values. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 88 New alarm or alert reinstates audible tone CLEAR Key: Clears selected numeric value. Audio Volume Key: Sets audio volume for alarms, alerts, and KVO tone. Numeric Keypad: Enters and changes values. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 89 POWER HOLD HOLD ml/hr ml/hr Flow Control Actuator KVO OPT HLD PRI SEC HLD OPT KVO OPTIONS Pumping Mechanism Clamp Arms Loading Guide • Clear Enter Pressure Transducer Air-in-Line Detector Air-in-Line Arm User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 90: Displays And Soft Keys

    Rate=125.0 Rate=130.5 VTBI= VTBI= Press key to switch immediately to split screen. DOPamine 10.0 460.0 40.0 Press key to stop split screen. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 91 The gauge updates for each program change during an infusion. Battery run time might be affected by the operating mode, rate, monitoring options, and back pressure. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 92 When the entry is complete, press the ENTER ENTER key. When all parameters required on a programming setup screen have been programmed, press the ok soft key to confirm all entries and continue programming. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 93: Configurable Settings

    The configurable options are set in the configuration mode by qualified service personnel. If the configuration settings need to be changed from the factory default settings, refer to the Technical Service Manual or contact CareFusion Technical Support for technical, troubleshooting, and preventive maintenance information.
  • Page 94 When Trends is enabled and the instrument is operating in pressure mode, a pressure trend graph is available. When Trends is enabled and the instrument is operating in resistance mode, a resistance trend graph is available. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 95 In Adjustable Pressure monitoring mode, the pressure alarm limit may be Pressure Limit, Maximum varied from 25 mmHg to this maximum configured pressure limit, in 25 mmHg increments. A value that exceeds this pressure limit cannot be selected. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 96 KVO rate or remains at the current rate, whichever is less. The KVO rate may be set from 0.1 mL/h to 20 mL/h in 0.1 mL/h increments and cannot be adjusted by the clinician. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 97 VTBI after each step of the infusion. The infusion may be programmed in rate/volume or volume/time. Rate, Maximum Maximum infusion rate may be set from 0.1 mL/h to 999.9 mL/h in 0.1 mL/h increments. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 98: Configurable Options

    On, Off Optional Features: Multi-Dose Alert On, Off Panel Lock On, Off VTBI On, Off Optional Modes: Dose Rate Calculator On, Off Loading Dose On, Off Multi-Dose On, Off Multi-Step On, Off User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 99 Resistance Alert On, Off Resistance Display On, Off Resistance Pressure Setting 1 to 600 mmHg 600 mmHg Rate, Maximum 0.1 to 999.9 mL/h 999.9 mL/h This feature is configured in the Diagnostics Mode. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 100: Specifications

    Specifications General Information Specifications Administration Sets Use only administration sets for Alaris SE pump. Alarms Accumulated Air In Line Key Stuck Air In Line Latch Open Battery Depleted No Upstream Flow Detected Channel Malfunction Occlusion Downstream Computer Link Failure Occlusion Upstream...
  • Page 101 500 to 1060 hPa Relative humidity 20 to 90% 5 to 95% Noncondensing Noncondensing Temperature Range 41 to 104°F -40 to 140°F (5 to 40°C) (-40 to 60°C) Fluid Ingress Protection IPX1, Drip Proof User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 102 VTBI Range 0.1 to 9999.9 mL in 0.1 mL increments (Basic Infusion, Drug Library Primary Infusion, IV Fluid Infusion, and Multi-Step mode); 0.1 to 999.9 mL in 0.1 mL increments (all other modes) User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 103: Symbols And Terms

    Flow sensor receptacle (optional), channel B. Infusing indicator. Indicates degree of protection, liquid ingress. IPX1 Manufacturing Date: Number adjacent to symbol indicates month and year of manufacture. MM-YYYY Nurse Call (optional for 7130/7230). Consult operating instructions. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 104 Drops per milliliter specification for product is identified on drop symbol. Product incorporates SmartSite needle-free valve ports and should not be accessed by a needle. Approximate set priming volume. ≈ XX ml User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 105 Depending on manufacturing and distribution timing, the Alaris SE pump might bear either the CF or CF Defibrillator-Proof symbol on the main rating label. The Alaris SE pump has been tested and complies with IEC 60601-1 Amendment 2, Clause 17 (h) for Defibrillator-Proof Equipment.
  • Page 106: Trumpet And Start-Up Curves

    Knowledge of the start-up characteristics should also be considered. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 107: Flow Characteristics Under Varying Delivery Conditions

    Under conditions of + mmHg pressure, the With a negative head height of - meters, the Alaris SE pump typically exhibits a long-term Alaris SE pump typically exhibits a long–term accuracy offset of approximately - from mean accuracy offset of approximately - from mean 1.4%...
  • Page 108: Pressure Mode

    Pressure Mode Trumpet Curve at 1 mL/h ( 48 hr ) Pressure Mode Trumpet Curve at 0.1 mL/h (48 hr) 20.3 10.7 98.0 51.0 23.8 12.3 -11.9 -7.5 -21.8 -6.2 -42.9 -13.9 -124.5 -100 Observation Interval (min) Observation Interval (min) User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 109 Pressure Mode Trumpet Curve at 25 mL/h (48 hr) -0.1 -0.1 -0.2 -0.3 -0.3 -0.7 -0.5 -3.3 -3.6 -3.4 -3.5 -3.5 -1.1 -3.7 -3.7 -3.6 -3.8 -3.6 -4.0 Observation Interval (min) Observation Interval (min) User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 110: Resistance Mode

    Resistance Mode Trumpet Curve at 1 mL/h (48 hr) 173.6 33.1 14.2 74.6 46.4 13.3 17.2 -1.5 -0.6 -1.0 -6.7 -22.0 -22.9 -48.9 -23.5 -109.5 -11.7 -100 Observation Interval (min) Observation Interval (min) User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 111 Resistance Mode Trumpet Curve at 999.9 mL/h (24 hr) 13.4 -0.2 -2.7 -2.9 -3.2 -0.5 -3.1 -3.1 -2.0 -3.3 -3.6 -3.5 -3.4 -3.7 -3.3 -3.9 -9.9 Observation Interval (min) Observation Interval (min) User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 112: High Resistance Mode

    -1.1 -1.4 -0.9 -1.3 -2.3 Observation Interval (min) High Resistance Mode Trumpet Curve at 999.9 mL/h (24 hr) -1.7 -2.7 -2.5 -3.0 -2.8 -3.1 -3.5 -3.9 -3.3 -3.7 -4.7 Observation Interval (min) User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 113: Troubleshooting And Maintenance

    To determine if abnormalities are caused by patient condition or the equipment, place the Alaris SE pump on hold. If the readings become normal, the equipment requires attention. Proper setup of the equipment should eliminate these artifacts.
  • Page 114: Air Bubble Detection

    • alarm tone sounds • rate LED display flashes • message appears in Main Display when channel is selected Opening the latch or turning the channel off clears air memory. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 115: Alarms

    Ensure air-in-line sensors are thoroughly cleaned. Battery Depleted (Plug In) Battery is too low to operate Plug power cord into an outlet instrument. immediately. Press run soft key or to resume infusion. HOLD User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 116 No Upstream Flow Detected alarm—small, rapidly falling drops form a continuous stream that does not trigger the flow sensor. In this event, unplug the flow sensor from the instrument. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 117 Press hold soft key to return to hold keys were pressed for ten mode. minutes. Instrument turns off if left in alarm more than five minutes. If an audio alarm remains on, turn instrument on and then off. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 118: Alerts

    Resistance Alert feature is on. Additional Alerts: Additional alerts provide notification of program completion and/or transition to another mode: Dose Complete (Multi-Dose mode), Loading Dose Complete, Multi-Step Complete, Secondary Complete, Infusion in VTBI User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 119: Prompts

    Occlusion Downstream A very high pressure exists in fluid Remove source of high pressure line while baseline is being set. and repeat setting of pressure baseline. Pressure Baseline feature is on. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 120 Reduce or temporarily remove Baseline due to motion, flow from other sources of variation while instruments, or blood pressure performing manual baseline setting prevents accurate setting of operation. pressure baseline. Pressure Baseline feature is on. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 121 During an infusion, if an invalid key is pressed, the following prompts might be seen: Dose Rate Running, Loading Dose Running, Multi-Dose Running, Multi-Step Running, Pri Running, Sec Running, or Timer Running (Multi-Dose program). User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 122: Inspection Requirements

    Inspect Cleaning As required (improperly lit/unlit pixels). Start-Up Each usage Cleaning Inspect and clean the Alaris SE pump as needed, typically WARNING between patient uses. Turn the instrument off and unplug the power cord from power use: DO NOT before cleaning.
  • Page 123: Service Information

    Qualified hospital/facility reference the return authorization information, and return it to personnel are responsible for the appropriate service or distribution center. CareFusion does checking in the instrument and not assume any responsibility for loss of, or damage to, ensuring that the current returned instruments while in transit.
  • Page 124 CareFusion warrants that: • Each new Alaris SE pump is free from defects in material and workmanship under normal use and service for a period of one ( ) year from the date of delivery by CareFusion to the original purchaser.
  • Page 125: Regulations And Standards

    The authority to operate this system is conditioned by the requirement that no modifications will be made to the system unless the changes or modifications are expressly approved by CareFusion. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 126 Cet appareil numerique de la Classe respecte toutes les exigences du Reglement sur le materiel brouilleur du Canada. Tables: The Alaris SE pump is intended for use in the electromagnetic environments specified in the following tables. Table 1 Electromagnetic Emissions...
  • Page 127 Power Frequency 400 A/m 60 Hz in a typical commercial or hospital Magnetic Field (50/ environment. 60 Hz) Compliance levels raised by IEC 60601-2-24 Performed at the minimum and maximum rated input voltage. User Manual—Alaris SE Pump, Rev2.X Models 7100/7130, 7200/7230...
  • Page 128 Voltage Dips, Short for 0.5 cycle for 0.5 cycle Interruptions, and If continued operation of the Alaris SE pump is required during power mains Voltage Variations 40% U 40% U interruptions, it is recommended that the...
  • Page 129 3 Vrms 10 Vrms Portable and mobile RF communications Conducted RF 150 kHz - 80 MHZ equipment should be used no closer to the Alaris SE pump (including cables) than IEC 61000-4-3 3V/m 10 V/m recommended separation distance Radiated RF 80 MHZ - 2.5 GHz...
  • Page 130 To assess the electromagnetic environment due to fixed transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Alaris SE pump is used exceeds the applicable compliance level, the Alaris SE pump should be observed to verify normal operation.
  • Page 131 Reduce the potential for electromagnetic interference by maintaining a minimum distance between portable and mobile communications equipment (transmitters) and the Alaris SE pump as recommended in this table, based on the maximum output power of the communications equipment. For transmitters rated at a maximum output power not listed in this table, the recommended separation...
  • Page 132: Standards

    Trademarks Regulations and Standards Compliance (Continued) Standards The Alaris SE pump has been assessed and complies with the following standards: , including amendments IEC EN 60601–1 BS 5724 , Group , Class Emissions; IEC EN 60601–2–24 CISPR 11 IEC EN 60601–1–2...
  • Page 134 User Manual—Alaris SE Pump, Rev2.X Models7100/7130, 7200/7230...

This manual is also suitable for:

7100713072007230

Table of Contents