Dynamic Monitoring System
Monitoring Options—Pressure Mode (Continued
Adjustable Pressure Alarm (Continued)
Automatic Baseline Calibration Used to
Monitor Pressure (Continued)
○ If current system pressure is same or higher than
original baseline, pressure baseline does not change.
○ If current system pressure is less than original baseline,
system automatically resets to new system pressure
value.
The pressure measurement can be optimized, particularly
at low flow rates (less than
restarting at least once every two hours (for example, when
reprogramming
VTBI
calibrate based on current system pressure.
Prior to activation, ensure that pressure has not built up in
the
line due to an occlusion or flow from other
IV
instruments through a common catheter. This results in a
more accurate pressure measurement.
When loading a set connected to a small diameter catheter,
wait at least five seconds after loading the set before
activating
/
RUN
HOLD
the loading process to dissipate and the sensor to stabilize.
Very small
PICC
might require
60
When multiple instruments are infusing through a common
small diameter catheter, pressure measurement accuracy
can be optimized by temporarily stopping all infusions, and
then restarting all instruments beginning with the instrument
delivering at the lowest rate.
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
mL/h), by pausing and
3
). This allows the pressure baseline to
. This allows pressure generated by
catheters, such as,
28
seconds or more for stabilization.
gauge/
French,
1.2
Monitoring Options—Pressure Mode
59