Prompts

Prompts alert the user but may not stop the infusion and are indicated by an audible sound, a flashing amber warning indicator and a
message on the display.
1. Check the display for a prompt message. Press
2. Rectify the cause of the prompt or proceed with caution.
Display
ATTENTION
SET VTBI
SET NOT FITTED
LOCKED
ADD DRUG
c
to silence the sound for 2 minutes, CANCEL to clear the message.
Infusion
Status
Pump on hold
Pump on hold
Pump on hold
Infusion continues
Infusion continues
1000DF00546 Issue 5
Alaris™ GP (Guardrails™) Volumetric Pump
Cause
Pump left on hold for 2
minutes without starting the
infusion.
No VTBI / flow sensor.
No infusion set fitted.
Rate change attempted
whilst locked.
Drug selection required.
31/50
Action
• Review pump setup.
• Start infusion or turn off pump.
• Set VTBI or fit flow sensor.
• Fit infusion set.
• Unlock rate to adjust infusion settings.
d
• Press
to access options menu.
• Select DRUGS A-Z using the
f
keys.
• Press OK to confirm.