Troubleshooting and Maintenance
Alarm
Occlusion Downstream
Occlusion Upstream
Primary Flow Detected During
Secondary
Set Out
Setup Time Exceeded
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
Alarms, Alerts, Prompts (Continued)
Alarms (Continued)
Pressure in
pressure alarm threshold.
Resistance has reached
Flow is obstructed between fluid
container and instrument's
pressure sensor.
Instrument detected flow from
primary container during
secondary infusion. (Alarm can
occur only when using an optional
flow sensor.)
Set was removed during an
infusion.
Instrument was turned on but no
keys were pressed for ten
minutes.
Meaning
line exceeded a
IV
.
100%
Alarms, Alerts, Prompts
Response
Check administration set for
probable cause (such as kinked
tubing, closed stopcock, high
resistance catheter). Press run soft
key to restart infusion.
Check administration set for
probable cause (such as kinked
tubing, closed clamp). Verify that
blue thumb clamp on Flow
Regulator has moved to open (up)
position. If not, reload set. Press
run soft key to restart infusion.
Verify that flow sensor is on Primary
line and that setup is correct.
Reinstall set. Press run soft key to
restart infusion
Press hold soft key to return to hold
mode.
Instrument turns off if left in alarm
more than five minutes.
If an audio alarm remains on, turn
instrument on and then off.