CareFusion Alaris SE User Manual page 95

Table of Contents

Advertisement

General Information
Definitions (Continued)
Dynamic Monitoring—Pressure
Adjustable pressure is designed to monitor
alarm limits. It is used for Precision Flow mode or for high resistance systems; such as, infusion through
transducers, into dialysis systems, and through highest resistance catheters. The pressure mode may
be configured to operate with a fixed pressure alarm limit threshold so that it is not user-adjustable.
Manual Pressure Baseline
Pressure Alarm
Pressure Display
Pressure Limit, Initial
Pressure Limit, Maximum
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
Configurable Settings (Continued)
Provides a real-time bar graph and numeric display of line pressure.
When Pressure Display is enabled, Pressure Alarm may be set to Adjustable
Pressure mode or Fixed Pressure mode.
In the Adjustable Pressure monitoring mode, the pressure alarm limit may be
varied from 25 mmHg to the maximum configured pressure limit, in 25 mmHg
increments. The pressure display indicates the current pressure limit and
provides the ability to adjust the limit by pressing the increase or decrease
arrows.
In the Fixed Pressure mode, the pressure limit of 600 mmHg is displayed, with
no means of adjusting it. When Pressure Display is disabled, the instrument
automatically defaults to Fixed Pressure mode.
Default pressure alarm limit that is automatically set when the instrument is
powered on. The alarm level must be less than or equal to the maximum
pressure limit.
In Adjustable Pressure monitoring mode, the pressure alarm limit may be
varied from 25 mmHg to this maximum configured pressure limit, in 25 mmHg
increments. A value that exceeds this pressure limit cannot be selected.
line/site pressure and provide user-adjustable pressure
IV
Configurable Settings
89