Prompts

Prompts alert the user but may not stop the infusion and are indicated by an audible sound, a flashing amber alert indicator and a
message on the display .
1 . Check the display for a prompt message . Press
2 . Rectify the cause of the prompt or proceed with caution .
Display
Cause
Pump left on hold for 2 minutes without starting
ATTENTION
the infusion .
SET VTBI
No VTBI/flow sensor .
SET NOT FITTED
No infusion set fitted .
LOCKED
Rate change attempted while locked .
ADD DRUG
Drug selection required .
c
to silence the sound for 2 minutes, CANCEL to clear the message .
1000DF00660 Issue 9
Alaris™ VP Plus Guardrails™ Volumetric Pump
Action
• Review Pump setup .
• Start infusion or turn off Pump .
• Set VTBI or fit flow sensor .
• Fit infusion set .
• Unlock rate to adjust infusion settings .
d
• Press
to access options menu .
• Select DRUGS A-Z using the
• Press OK to confirm .
32/48
Prompts
f
keys .