WARNING
When priming:
○ Ensure that patient is not connected.
○ Ensure that air is expelled from line prior to beginning
infusion (unexpelled air in line could have serious
consequences).
Failure to prime correctly can delay infusion delivery and
cause the total volume to be infused to read higher than the
actual total delivered to the patient.
Discard if packaging is not intact or protector caps are
unattached.
Use only sets dedicated for use with the Alaris SE pump.
The use of any other set might cause improper instrument
operation, resulting in an inaccurate fluid delivery or other
potential hazard.
CAUTION
Before operating the instrument, verify that the administration
set is free from kinks and installed correctly in the
instrument.

Epidural Administration

WARNING
Epidural administration of drugs other than those
indicated for epidural use might result in serious injury to
the patient.
It is strongly recommended that the source container,
administration set, and Alaris SE pump used for epidural
drug delivery be clearly differentiated from those used for
other types of administration.
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
Warnings and Cautions (Continued)

Administration Sets

Warnings and Cautions