Adjustable Pressure Alarm - CareFusion Alaris SE User Manual

Table of Contents

Advertisement

Dynamic Monitoring System
Monitoring Options—Pressure Mode (Continued
Auto Restart Plus Feature (Continued)
This feature can be configured through the hospital
configuration settings to allow from one to nine Checking Line
restarts. After the programmed number of restarts has
occurred or the
-second Checking Line period has been
40
exceeded, the channel immediately alarms
when resistance or pressure conditions
DOWNSTREAM
indicate an occlusion.

Adjustable Pressure Alarm

In the Adjustable Pressure monitoring mode, the pressure
alarm limit may be varied from
configured pressure limit, in
alarm level and a maximum pressure limit are set through the
hospital configuration settings.
Set Alarm Limit Marker
Pressing the
or
mmHg in the corresponding direction. It is advisable to
25
select an alarm limit appropriate for the flow rate. At lower flow
rates, the alarm limit should be set lower, to shorten the time to
alarm.
Automatic Baseline Calibration Used to
Monitor Pressure
The auto pressure baseline calibration remains in effect until
the instrument is turned off, the latch is opened, the set is
reloaded, or the Set Pressure Baseline function is performed.
The first activation of
automatically establishes a pressure baseline based on
current system pressure. An optimal baseline is
maintained upon subsequent activations of
follows:
58
mmHg to the maximum
25
mmHg increments. A default
25
soft key changes the alarm limit by
for a new infusion
/
RUN
HOLD
OCCLUSION
, as
/
RUN
HOLD
Monitoring Options—Pressure Mode
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230