Dynamic Monitoring System
Monitoring Options—Pressure Mode (Continued)
\
Pressure Baseline (Continued)
Manually Set Pressure Baseline While
Operating in Adjustable Pressure Mode
(Continued)
4.
Press Set Pressure Baseline soft key.
Set Pressure Baseline screen appears.
5.
Verify that no pressure, due to occlusion or other infusions
through a common line, is present in
6.
For best results, verify that set outlet (for example,
stopcock) is located at patient's heart level.
7.
Press ok soft key.
8.
Verify that pressure readout is
1.
To return to the normal screen without setting the baseline, press
the return soft key.
2.
True baseline pressure will be zero or within a few mmHg of zero.
If not, and the pressure is unstable, allow the pressure to drop to
the lowest level and then repeat the Set Pressure Baseline
process.
User Manual—Alaris SE Pump, Rev2.X
Models 7100/7130, 7200/7230
1
line.
IV
A
B
Set
Pressure
Baseline
return
(zero) mmHg.
2
0
Monitoring Options—Pressure Mode
OPTIONS
A B
A
Rate=125.0 mL/h
VTBI=500.0 mL
ok
A B
Rate=
VTBI= 1000.0 mL
VI=
OPTIONS
A
New
Baseline
Set
OPTIONS
125.0 mL/h
A
0.0 mL
Primary
0 mmHg
61